Use the Back button in your browser to see the other results of your search or to select another record.
Nasal jet-CPAP (variable flow) versus bubble-CPAP in preterm infants with respiratory distress: an open label, randomized controlled trial |
Bhatti A, Khan J, Murki S, Sundaram V, Saini SS, Kumar P |
Journal of Perinatology 2015 Nov;35(11):935-940 |
clinical trial |
6/10 [Eligibility criteria: Yes; Random allocation: Yes; Concealed allocation: Yes; Baseline comparability: No; Blind subjects: No; Blind therapists: No; Blind assessors: No; Adequate follow-up: Yes; Intention-to-treat analysis: Yes; Between-group comparisons: Yes; Point estimates and variability: Yes. Note: Eligibility criteria item does not contribute to total score] *This score has been confirmed* |
OBJECTIVE: To compare the failure rates between Jet continuous positive airway pressure device (J-CPAP-variable flow) and bubble continuous positive airway device (B-CPAP) in preterm infants with respiratory distress. STUDY DESIGN: Preterm newborns < 34 weeks gestation with onset of respiratory distress within 6 h of life were randomized to receive J-CPAP (a variable flow device) or B-CPAP (continuous flow device). A standardized protocol was followed for titration, weaning and removal of CPAP. Pressure was monitored close to the nares in both the devices every 6 hours and settings were adjusted to provide desired CPAP. The primary outcome was CPAP failure rate within 72 h of life. Secondary outcomes were CPAP failure within 7 days of life, need for surfactant post-randomization, time to CPAP failure, duration of CPAP and complications of prematurity. An intention to treat analysis was done. RESULTS: One-hundred seventy neonates were randomized, 80 to J-CPAP and 90 to B-CPAP. CPAP failure rates within 72 h were similar in infants who received J-CPAP and in those who received B-CPAP (29 versus 21%; relative risks 1.4 (0.8 to 2.3), p = 0.25). Mean (95% confidence intervals) time to CPAP failure was 59 h (54 to 64) in the Jet CPAP group in comparison with 65 h (62 to 68) in the bubble CPAP group (log rank p = 0.19). All other secondary outcomes were similar between the two groups. CONCLUSION: In preterm infants with respiratory distress starting within 6 h of life, CPAP failure rates were similar with jet CPAP and bubble CPAP.
|